|

FLOW CYTOMETRY CORE FACILITY
INTRODUCTION
Flow
Cytometry Facility
The Flow Cytometry Facility is a shared resource, which is available
to all Cornell University researchers, and others. We are currently
operational with two Beckman-Coulter XL flow cytometers, an Altra
cell sorter, and a CompuCyte Laser Scanning Cytometer (LSC). The
XL is a streamlined, user-friendly instrument and can be learned
quite easily in a two-hour training course which is required for
new users who wish to operate the instrument. Christopher Colon,
the facility's Director, will be training and assisting all users.
The Director is available to assist researchers in the development
of flow cytometric assays and protocols for the preparation of samples,
although it is required that all sample preparation be performed
at the researcher's own facility unless otherwise arranged.
The
facility is located at Weill Medical College at 1300 York Avenue,
Room A-1027, (212) 746-6782.
Independent
use
Users
who have operated the XL analyzer instruments will be allowed
unassisted use upon evidence of proficiency. Training in the use
of the instruments is available on either an individual or group
basis for a fee (see fee schedule) after which a user may be able
to operate the instrument independently. The Altra cell sorter
will be operated by core facility personnel only, with the requirement
that the investigator be present for Altra sorting.
Assisted/
supervised use
The facility's staff or the Director will assist investigators
in using the instrument based on availability, and will provide
information for the preparation of samples for cytometric analysis.
Samples must be prepared at the investigator's lab and must
be in good condition prior to loading onto the instrument. Guidelines
for sample preparation will be discussed prior to preparation,
and well before presentation of the samples to the cytometers.
If it becomes necessary to fully assist the user until s/he
is able to be completely independent, then the charges that
will apply are those listed for "Assisted Use" in
the fee schedule below. The Facility is minimally staffed and
does not accept samples for processing as a reference lab except
by prior arrangement with the Director (LSC sample analysis
may be an example of such an arrangement if the sample is in
optimal condition; prepayment is required).
Assay/
protocol development
Flow cytometric analysis may not be appropriate for an existing
preparation, in which case it will be necessary to develop a
novel method of preparation or to research the literature for
such a method. In these instances, the Director will be available
for collaboration with the investigator. It is helpful for any
investigator who wishes this type of collaboration to familiarize
him or herself as much as possible with flow cytometric analysis
and techniques prior to meeting with the Director. The publication,
"Current Protocols in Cytometry" will be available on CD in
the facility as a reference for assay development in flow cytometry.
There are references for LSC sample preparation and analysis
as well, however, this form of cytometric analysis being relatively
new, may require that the investigator develop a unique sample
preparation protocol.
Facility
Description
The Flow Cytometry and Cell Sorting Facility is located
at Weill Medical College, Room A-1027. The facility also has
a Citrix server available for Cornell scientists who generate
high volume data for analysis. Software is available for DNA
cell cycle analysis, listmode analysis, and graphic display
of data generated by most commerical flow cytometers (FCS format).
INSTRUMENTATION
Two
Coulter XL analytical flow cytometers.
The
XL is a Beckman-Coulter productwhich is designed for analytical
flow cytometry and is easily operated with some basic instruction.
The system features the capacity to analyze simultaneously up
to four fluorescence wavelengths from fluorochromes which are
excited by the instrument's air cooled argon laser (at 488 nm).
Laser
Scanning Cytometer (LSC).
The Laser Scanning Cytometer (LSC) from CompuCyte Corporation
is the latest instrument in the flow cytometry core facility.
This instrument is used to measure fluorescence characteristics
of cells on regular glass microscope slides, or their equivalent
(75mm x 25mm x 1mm; read below for more information). Cells scanned
with lasers are also available for visual inspection, unlike a
flow cytometer. Less sampling is required than for flow cytometry
and stabilized slides may be read repeatedly. The instrument has
two air cooled lasers, an Argon and a HeNe. A violet laser will
be added by the end of 2001.
Beckman-Coulter
Altra cell sorter.
The
Altra is marketed for the researcher who has the need to sort
cell sub-populations. It is equipped with various lasers for the
excitation of a wide variety of fluorochromes such as FITC, PE,
PI, Texas Red, APC. The
Altra is capable of sorting 10,000 events per second. This flow
cytometer is a flexible and powerful tool intended for researchers
who need the power and versatility not available with the analyzer
model
above.
ORGANIZATION
& SERVICES PROVIDED
Assay
Development/Protocol Design
The investigator may ask for the Director's assistance in the
development of novel assays/ protocols for a specific application,
in which case a fee will be charged.
Flow
Cytometric Analysis
Investigators
are expected to have met with the Director of the facility prior
to bringing any samples so that there will be as little sample loss
as possible. There is no provision for the analysis of radiolabeled
cells, or cells which have been physically treated with radioisotopes
and which pose a potential radiation safety hazard. Also, preparations
of potentially infectious materials must be fixed prior to analysis
to reduce the risk of contamination. Cells and particles within
the instrument specifications may be analyzed in concentrations
of no less than 500,000 cells or particles per milliliter of suspension
media. Typical analytes are mammalian cells at 1x10EE6 cells per
ml. Other particles such as algae, microbeads, and aggregates or
colloidal suspensions may be analyzed as long as they satisfy size
limits (2-150 microns depending on other characteristics such as
fluorescence intensity).
Laser
Scanning Cytometry
The Laser Scanning Cytometer (LSC) from CompuCyte Corporation
is the latest instrument in the flow cytometry core facility. This
instrument is used to measure fluorescence characteristics of cells
on regular microscope slides, or their equivalent (75mm x 25mm x
1 mm). The sample requirements are as follows (please call the director
for more information prior to any sample submissions):
-
Cells
must be free of one another and must be adhered to the surface
of a regular, plain, glass microscope slide (or other optically
clear, perfectly flat, non fluorescent, similarly sized material).
Your cells cannot be moving!
- Cells
should be in a concentration which would allow for a mono-layer
to be produced which will provide non-overlapping cells very close
to one another, but not touching one another. There should be
a stable fluorescent dye or fluorochrome present since this is
a fluorescence-based technology (excitation wavelengths are from
two lasers: Argon, HeNe).
Currently,
the LSC instrument requires single-cell type preparations, not tissue
sections (Compucyte has not fully developed its plan for this type
of sample preparation).
This
is an exciting and relatively new microscope slide-based technology
which may require you to develop a new way of sample preparation
for your own particular specimens. However, it is also one which
would potentially grant you greater flexibility in terms of sample
analysis than other methods currently available.
Because
the cells are adhered to the slide and cannot move around, the LSC
is able to record the position on the slide for each cell analyzed
for possible later re-staining and re-analysis. The LSC provides
data similar to flow cytometry data and which may be similarly manipulated.
The advantage over flow cytometry is that one does not need as many
cells, and that a sample may be analyzed more than once (data files
may be combined for discrete runs of the same slide).
The
following are examples of inappropriate samples:
- Slides
which are either smaller or larger than standard microscope slides
(standard slides are 75mm x 25mm x 1mm).
- Cells
which are already fixed cannot be used (fixed cells do not stick
or adhere to slides so that if you are going to fix your cells
for staining it is best to do your staining on live cells, then
allow cells to adhere to the slide for further treatment such
as fixation and additional staining).
- Cells
which are overlapping and are crowded on the slide (cells in very
low concentrations take longer to read and are difficult to set
up. If you have a shortage of cells, you should attempt to concentrate
them onto a smaller area on the slide using a cytospin prep if
necessary, but remembering not to crowd them either!)
Training
Training
in the use of the instruments is available to Cornell University
users and for those who wish to familiarize themselves on the use
and application of flow cytometry. This is a basic course that is
intended for investigators who wish to operate unassisted. Also,
basic instruction on software is available, although it is expected
that each user will familiarize themselves with the software and
will be analyzing data independently.
HOW
DO I GET STARTED?
Contact
the Director of the Flow Core Facility
Christopher
Colon, 212-746-6782
Schedule
a session
A calendar is available online for convenient scheduling upon which
you must place your name, the desired instrument, your phone/extension,
and time requirement, at least 24 hours in advance. Please contact
the facility to arrange for your first appointment and to obtain
the URL for the online calendar, as well as a username and password
for future appointments.
Cancellation
policy
XL
and XLMCL analyzers:
You must cancel your reservation at least 24 hours prior to your
reserved time by placing "CANCELLED" in front of your
original reservation on the calendar. If you cancel more than
24 hours in advance you may delete your reservation completely
from the calendar. Your account will be billed for the reservation
unless you cancel at least 24 hours in advance in order to be
fair to everyone, and to allow everyone to have equal access to
the instruments (there are no exceptions; however, please feel
free to come to the director if you feel you should be exempt
from this policy). If the instrument is not functioning, you will
not be charged for unused time.
Sorter
and LSC:
You
must cancel your reservation for the sorter or the LSC at least
72 hours in advance to avoid a charge, and to allow others to
have ample opportunity to utilize the available time. The charge
for cancellation will be a maximum of $125.00 regardless of the
cause (avoidable, or unavoidable). As always, you will not be
charged if another user makes use of your cancelled reservation.
Please
note: If you keep staff over closing time during your session you
will be charged at the assisted rate for the entire session. Please
be on time and prepare to stop your analysis at least 10-15 minutes
prior to your end time so that you have enough time to remove your
data and can complete the instrument cleaning protocol.
Billing
is as stated in the fee schedule.
ALL users must provide a valid account number prior to utilization
of the facility.
| Instrument/Service |
Cornell Rate |
Outside Rate |
| Assisted |
Unassisted |
Assisted |
Unassisted |
| Altra |
75 |
N/A |
150 |
N/A |
| XL |
75 |
35 |
150 |
N/A |
| Data Archival Disks |
20 |
20 |
| Photo Quality Prints |
10 |
20 |
| Sorting |
75 |
150 |
| Laser Scanning Cytometer |
125 |
90 |
250 |
N/A |
| Applications support/Assay development |
125 |
250 |
| Flow Cytometry Training Session |
275 |
275 |
Rates/charges
are per hour or any part of an hour. The facility reserves the right
to change rates at any time without notice.
Notes:
1.
Set-up time for sorter is charged at the hourly rate for Altra use
(advanced notice IS required for complex set-ups).
2.
All Altra and LSC operation is by Core Facility personnel.
3.
Charge will apply if user does not supply disk.
4.
Set-up cost for "clean-sort" will be $150 (advanced notice IS
required).
5.
Assistance is available for development of flow cytometric assays
and procedures upon request, and prior arrangement (any reagent
or incidental costs are the responsibility of the user).
Related
Web Links

© Weill Medical College of Cornell University. All rights reserved. | Privacy Notice
|